Thursday, 28 May 2009

10.standard procedure in an electrical driven machine

Laboratory supply of mains electricity is via individual socket outlets that may be 3-phase 415 V, 50 Hz, or single-phase 240 V, 50 Hz. Office supplies are normally 240 V, 50 Hz. Most electrical equipment operates on 240 V, 50 Hz.A wide range of electrically powered equipment is found in the laboratory including fluid and vacuum pumps, lasers, power supplies, electrophoresis and electrochemical apparatus, X-ray equipment, stirrers, hot plates, ovens both conventional and microwave and computers, printers and VDU equipment. In the office, there are, amongst other things, computer equipment, FAX machines and photocopiers. In the lecture theatres there are overhead and slide projectors. In fact, everyone in the School of Chemistry is exposed at some time to electrical equipment.


Hazards
· Electric shock is the effect produced on the body and particularly on the nervous system by an electrical current passing through it. The effect depends on the current strength which itself depends on the voltage and body resistance i.e. path length and surface resistance of skin (which is much reduced when wet). Death can be the result of the normal voltage of 240 V causing currents of greater than 30 mA to flow through the body for more than 40 ms. Minor shocks may also cause injury following involuntary muscle contraction.
· Burns caused by the passage of heavy currents through the body or by direct contact with an electrically heated surface.
· Explosion and fire caused by electrical sparks, short circuits or overload heating, old wiring in the presence of flammable material. Injury from microwave and radio-frequency sources and from induction heating equipment.
Control Measures
These precautions are not meant to be exhaustive or to cover aspects of repair or construction of electrical equipment (details of such can be found in the University Code of Practice "Electrical Safety") but to cover everyday use in the laboratory and office.
Plugs and fuses
· Plugs that are cracked or broken must not be used. The plugs must be wired properly, the conductors securely fixed and the cable firmly held by the strain relief grip.
· The rating of the fuse must be appropriate to the appliance. Most electronic equipment (computers, measuring instruments etc.) requires only a 3 A fuse which will load to 720 W. Reserve 13 A fuses (loading to 3000 W) for heavier equipment. Electronic Workshop personnel will advise in case of doubt.
Cabling
· The cable must be in good condition and free from breaks in the insulation. Cable must be sufficiently robust to withstand the wear and tear of laboratory or office use and fully waterproof where water may come within the vicinity of the apparatus.
· Cables must not be run across the floor in such a way as to cause a tripping hazard or to be susceptible to damage from passing traffic. If it is necessary to run cables across walkways, they must be covered with cable protectors.
Extensions /Adaptors
It is permissible if necessary to feed one four way extension block from a single socket provided the block feeds only low power equipment (less than 500 W or 2 A). Extension leads must not be daisy-chained. Kettles, microwaves and heaters that have higher power demands must not be used on such an extension but must be fed from an installed socket point.
240V multi-way adaptors are forbidden by the University
Mains Switch
The location of any mains switch must be clear and known so that power can be turned off rapidly in an emergency.
Use
· No apparatus with exposed mains terminals should ever be used.
· Ordinary electrical equipment must not be used in the vicinity of flammable or explosive gases. Ordinary electrical equipment is a possible source of ignition.
· Ordinary electrical equipment must not be used where it may get wet. Water may cause a dangerous short circuit.
· Equipment that has been wet must never be switched on until the equipment has been tested. Anyone to whom the equipment is taken for testing must be informed about what has happened.
Repairs
· Electrical equipment must not be "repaired" except by a competent person.
· Equipment must be disconnected from the main power before beginning.
· If there is any doubt, equipment should be taken to the Electronics Workshop for repair.
Testing
It is required by law that electrical equipment is tested from time to time. The School of Chemistry policy on this matter is described here. Most portable items of equipment i.e. those that can be unplugged from the wall, carry a School of Biological and Chemical Sciences identification label and a most recent testing date. Apparatus with a FAILED sticker must never be used. Before embarking on more sophisticated electrical work such as building custom equipment, the University Code of Practice entitled "Electrical Safety" should be consulted with particular attention paid to the regulations covering the proper insulation of conductors and